| Mode | CPAP, Auto CPAP |
| Pressures delivered | 4-20 cm H20 (± 0.5 cm H2O), adjustable in steps of 0.5 cm H2O |
| Max. flow rate for a pressure drop of 1 cm H2O at patient inlet | 150 I/min at 20 cm H2O without humidifier |
| Ramp time | 0-45 minutes, adjustable in 5 minutes steps |
| Noise level | 27.5 dBA measured with a tolerance of 3 dBA according to NF EN ISO 17510-1:2009 |
| Size | Width 14 cm x Length 20 cm x Height 8 cm |
| Weight | 0.750 kg without power supply |
| Electrical supply | Input range 100-240V (50-60 Hz) |
| Maximum electrical power | 25 W (without humidifier) |
| 93 / 42 / EC Classification | Class II A |
| Temperature | +5 °C to +40 °C, +5 °C to +35 °C with humidifier |
| Humidity | 10 – 95% relative humidity without condensation |
| Storage | –20 °C to +60 °C, ± 95% relative humidity without condensation |
| Warranty | Three years |
| Altitude | From sea level to 2400 metres |